Automatic B-lines: a tool for minimizing time to diuretic administration in pulmonary edema patients in the emergency department of a developing country
International Journal of Emergency Medicine volume 17, Article number: 183 (2024)
Abstract
Background
Effective management of pulmonary edema in the emergency department (ED) is crucial given its significant global impact on health. This study aimed to investigate the hypothesis: “Does the utilization of Automatic B-lines via ultrasonography in patients with pulmonary edema facilitate faster diuretic administration in a developing country?”
Methods
This retrospective observational study was conducted at a tertiary academic center in Thailand. Patients with pulmonary edema admitted to the ED between January 2023 and June 2024 were enrolled. Ultrasound documentation and electronic ED medical records were compared to assess the time of diuretic administration between patients who had lung ultrasounds utilizing automatic B-lines and those who had manual B-lines counted by physician eye inspection. Multivariate logistic regression was employed to examine the relationship between the use of automatic B-lines and early diuretic administration.
Results
The study included 134 patients with pulmonary edema. The time to diuretic administration was significantly shorter in the automatic B-lines group (median time [Q1-Q3], 55 min; range, 35–110 min) compared to the non-automatic B-lines group (median time, 100 min; range, 75–145 min). In the multivariable logistic regression analysis, early diuretic administration within 60 min of triage was significantly more likely in the automatic B-lines group (adjusted odds ratio, 1.45; 95% confidence interval, 1.10–2.45) than in the non-automatic B-lines group.
Conclusions
In a developing country, patients with pulmonary edema who had lung ultrasound evaluation with automated B lines experienced a fastest diuresis compared to those who utilized ultrasonography without automatic B lines. Implementing automatic B-lines as an early screening protocol could enhance clinical practice in the ED.
Background
Pulmonary edema is a major public health concern that consumes a significant amount of healthcare resources worldwide. It is a leading cause of hospitalization among patients admitted to the emergency department (ED). Patients with pulmonary edema experience high mortality rates and an elevated incidence of rehospitalization particularly in developing country. As admissions for pulmonary edema increase, effective assessment and management in the ED are critical [1,2,3,4].
Lung ultrasound has been widely used as an additional diagnostic method for lung diseases [5,6,7]. It is a noninvasive tool used to differentiate between various lung pathologies, with diagnostic efficacy comparable to or even higher than that of other methods.
B-lines are one of the ultrasound artifacts that originate at the pleural line, appears as independent laser-like vertical hyperechoic reverberation artifacts that obscures other background lung ultrasound artifacts, which radiate to the bottom of the screen without fading and move continuously with lung sliding [8]. B-lines visible on lung ultrasound correlate with extravascular lung water content, aiding in the diagnosis of pulmonary edema [9,10,11,12]. However, ultrasonography is significantly operator-dependent, and differentiating B-lines from other vertical artifacts in lung ultrasonography can be particularly challenging, especially in developing countries where point-of-care ultrasound (POCUS) has not been widely utilized. Artificial intelligence contributes to various functions in healthcare [13,14,15]. Implementing reliable artificial intelligence real-time automatic tools, including automatic B-lines detection, can reduce testing variation and improve lung ultrasound examination interpretation, eliminating the need for manual assessment or measurements. Reducing reliance on operators’ abilities can make the information obtained more reliable and suitable for decision-making [16,17,18,19].
Following a diagnosis of pulmonary edema, early decongestion therapy with intravenous diuretics is critical, as it has been linked to better outcomes in pulmonary edema [20,21,22]. We hypothesized that the use of automatic B-lines detection in lung ultrasound for patients with suspected pulmonary edema would assist emergency physicians in developing immediate management strategies.
This study aimed to determine the relationship between the presence of automatic B-lines during lung ultrasound and the administration time of intravenous diuretics in patients admitted to the ED with pulmonary edema.
Methods
Design and setting
This retrospective observational study investigated the electronic medical records of patients with pulmonary edema who presented to the ED of a tertiary academic teaching hospital in Khon Kaen, Thailand, which serves approximately 60,000–70,000 emergency patients annually.
Patients
The participants in this study were patients who visited our ED between January 2023 and June 2024. Adult patients (age > 18 years) with pulmonary edema, as defined by the ICD-10 International Classification of Diseases, 10th edition, who presented to the ED and received intravenous furosemide were included. The exclusion criteria for our study included: 1) patients without ultrasound documentation, indicating incomplete records; those who did not undergo ultrasound; and individuals unable to interpret results; 2) patients lacking information on the timing of furosemide administration; and 3) patients transferred to another institution.
Data collection
The principal investigator reviewed electronic medical records at Srinagarind Hospital, Faculty of Medicine, Khon Kaen University, Thailand, using the Health Object Program®, an authorized electronic medical records system, for all patients with pulmonary edema who presented to the ED. Eligibility was determined based on the final diagnosis of pulmonary edema, as identified by the ICD-10 International Classification of Diseases, Code J81.0. Data were extracted from various sections of the electronic medical records, with each patient assigned an anonymous identity. The extracted data included demographic information from the patient’s profile sheet, medication administration details, nursing notes (including the time of furosemide administration), and ultrasound interpretation.
The ultrasound interpretations were compiled by emergency medicine residents and emergency physicians who routinely evaluate patients suspected of having pulmonary edema using POCUS, which includes ultrasound characteristics of the heart, lungs, and inferior vena cava. However, the decision to perform POCUS was dependent on the attending physician at the time. B-line artifacts in lung ultrasonography could be assessed using either automated B-lines or the physician’s visual inspection. The attending physician determined which B-line assessment approach to apply. The ultrasound machine’s automatic B-lines function, known as “Smart B-lines,” is an artificially intelligent feature that automatically detects and analyzes B-lines (Fig. 1).
This study utilized the Mindray M9 ultrasound machine (Mindray, Shenzhen, China). In terms of lung ultrasound examination, our study was carried out using the BLUE protocol, which included three standardized points: the upper BLUE point, the lower BLUE point, and the PLAPS point. The preset was set as lung preset. The transducer in this study was The Mindray C5-1s Curved Array transducer.
This study included emergency physicians as well as emergency medicine residents (postgraduate years 1–3). All these individuals had to attend a one-hour demonstration of Automatic B-lines. Throughout the academic year, emergency medicine residents completed a one-month ultrasound rotation.
The time to diuretic administration was defined as the interval between the patient’s arrival at the ED and the first intravenous administration of furosemide (as recorded in the nursing notes). The arrival time was defined as the time of the patient’s first medical contact, typically with a triage nurse (as noted on the profile sheet). In this study, early administration of intravenous diuretics was defined as within 60 min of triage, following the European Society of Cardiology’s recommendations for the early management of acute cardiogenic pulmonary edema [23].
Data were gathered and compiled into a research dataset. Two independent investigators then reviewed the data, and any redundant entries were addressed. If discrepancies were found, senior investigators were consulted to ensure the accuracy of the data.
The main objective of this study was to examine the relationship between the use of automatic B-lines during lung ultrasound and the timing of diuretic administration in patients admitted to the ED with pulmonary edema.
Analysis and statistics
A power analysis for a paired sample t-test was conducted to determine the minimum sample size, resulting in 66 patients required for each group [24]. The enrollment ratio was 1.0, with an alpha level of 0.14 and 80% power (β = 0.1), as estimated from a previous study [25].
Data were entered into Microsoft Excel (Microsoft Windows 10, Khon Kaen University license) and analyzed using IBM SPSS for Windows version 27.0, licensed by Khon Kaen University (SPSS Inc., Chicago, Illinois, USA).
Descriptive statistics included frequencies, means for normally distributed data, and medians (interquartile range [IQR]) for non-normally distributed data. To assess the significance of time to diuretic administration, a t-test or median test, chi-square test, and univariate/multivariate logistic regression models were utilized.
After univariate logistic regression, the principal investigator selects the variables with a p-value less than 0.20, and those variables proceed to multivariate logistic regression.
Results
Baseline characteristics
Between January 2023 and June 2024, 1,110 patients diagnosed with pulmonary edema using the ICD-10 International Classification of Diseases visited the ED. After excluding patients with missing ultrasound documentation (n = 750), missing furosemide administration times (n = 125), and those transferred to another institution (n = 101), 134 patients were included in the final evaluation. Automatic B-lines detection during lung ultrasound was performed in 67 patients (Fig. 2). Table 1 compares the individual characteristics between the automatic and non-automatic B-lines groups.
Outcome
According to the data presented in Table 2, which analyzed the time of furosemide administration between patients who underwent lung ultrasound using automatic B-lines and those using non-automatic B-lines, the time to diuretic administration was significantly shorter in the automatic B-lines group (median time [Q1-Q3], 55 min; range, 35–110 min) compared to the non-automatic B-lines group (median time, 100 min; range, 75–145 min).
In multiple logistic regression models adjusting for age, sex, mode of arrival, vital signs, medical history, diffuse B-lines in lung ultrasound, and plethoric inferior vena cava, early administration of diuretics within 60 min of triage was significantly more likely in the automatic B-lines group (adjusted odds ratio, 1.45; 95% confidence interval, 1.10–2.45) than in the non-automatic B-lines group.
Discussion
Pulmonary edema is a serious public health concern that causes significant morbidity, mortality, and economic costs. Recent studies have highlighted the critical role of the ED in managing pulmonary edema. While early diagnosis is essential, the effectiveness of treatment is often time-dependent [3, 26, 27]. Most studies have concluded that B-lines, an ultrasound artifact seen in lung ultrasound, are a valuable bedside tool for diagnosing pulmonary edema, aiding in diagnosis, monitoring, and prognosis. Persistent pulmonary congestion significantly increases the risk of rehospitalization [28,29,30]. However, in the setting of developing country, untrained sonographers may misdiagnose pathological interstitial syndrome by failing to distinguish B-lines from other vertical ultrasound artifacts [31]. To address this issue, an innovative feature known as the automatic B-lines function, which utilizes artificial intelligence in ultrasound machines, has been developed. This technology can enhance diagnostic efficiency by reducing operator dependency [32]. With the accurate results provided by automatic B-lines in lung ultrasound, emergency physicians can deliver prompt treatment, such as administering intravenous diuretics, which have been shown to reduce pulmonary congestion in cases of pulmonary edema.
This study investigates the relationship between diuretic administration time and the use of automatic B-lines during lung ultrasound in patients presenting with pulmonary edema in the ED. Our findings indicated that among patients with pulmonary edema undergoing ultrasound evaluation, those exhibiting automated B lines received diuresis administration more rapidly than those who utilized ultrasonography without automatic B lines. The median time was 55 min, which is less than half the median time in the non-automatic B-lines group (100 min). These results are consistent with previous research [25], which established that using POCUS, particularly cardiac ultrasound, as a bedside tool reduced the time to furosemide administration compared to the non-focused cardiac ultrasound group. Furthermore, a previous study found that using POCUS in the ED reduced the ED length of stay [33].
The results of the multivariable linear regression analyses indicated that the group using automatic B-lines had significantly higher rates of early diuretic administration within 60 min of triage (adjusted odds ratio, 1.45). This finding underscores that automatic B-lines during lung ultrasound can reduce the time required to administer intravenous diuretics. This is consistent with a previous study that found early administration of furosemide within 2 h of triage was significantly higher in the focused cardiac ultrasound group (adjusted odds ratio, 1.63). Implementing automatic B-lines in lung ultrasound for patients with pulmonary edema could enhance ED practices in the setting of developing country [25].
Strengths and limitations
One strength of our study is that, to our knowledge, we are the first to use automatic B-lines, a function in an ultrasonography machine that can distinguish B-lines from other lung ultrasound artifacts. Our study focused on using automatic B-lines to enhance acute pulmonary edema management in developing countries’ emergency departments. Our study had the following limitations: 1) It was conducted at a single tertiary academic teaching hospital, so the findings may only be representative of similar healthcare settings. 2) Only 134 of the 1,110 patients were included in this study, possibly as a result of the retrospective observational design, which may include biases in patient selection. The majority of the exclusions (750 patients, 67.56%) were related to the lack of ultrasound documentation, insufficient data, and confounding variables that could influence results. 3) The study focused on the immediate effect of automatic B-lines on diuretic administration. Further studies are needed to assess long-term outcomes, such as hospital length of stay and mortality rates. 4) The study utilized a specific ultrasound machine with automatic B-line functionality, so the results may not be generalizable to other ultrasound machines with different capabilities. 5) The study did not demonstrate the diuretic dose administration in patients and the correlation between diuretic dose and the number of B lines assessed. 6) The study did not examine the knowledge and experience levels of POCUS practitioners, which could influence the accuracy of assessing B-lines artifacts. However, the research aims to evaluate the effectiveness of automatic B-lines detection, a feature that could assist both experienced and novice physicians in making decisions about administering diuretics to patients with pulmonary edema.
Conclusion
Patients with pulmonary edema who utilized ultrasonography and exhibited automatic B-lines experienced faster diuresis administration than those who employed ultrasound without automatic B- lines in the setting of developing country. Early diuretic administration within 60 min of triage is significantly more common in the group utilizing automatic B-lines. Our findings suggest that emergency physicians can use automatic B-lines at the bedside to enhance the diagnosis and early management of pulmonary edema.
Data availability
No datasets were generated or analysed during the current study.
Abbreviations
- ED:
-
Emergency department
- POCUS:
-
Point-of-care ultrasound
- ICD-10:
-
International classification of disease, 10th edition
- IQR:
-
Interquartile range
- ESI:
-
Emergency severity index
References
Deniau B, Costanzo MR, Sliwa K, Asakage A, Mullens W, Mebazaa A. Acute heart failure: current pharmacological treatment and perspectives. Eur Heart J. 2023;44(44):4634–49.
Guttikonda SNR, Vadapalli K. Approach to undifferentiated dyspnea in emergency department: aids in rapid clinical decision-making. Int J Emerg Med. 2018;11(1):21.
Zanza C, Saglietti F, Tesauro M, Longhitano Y, Savioli G, Balzanelli MG, et al. Cardiogenic pulmonary edema in Emergency Medicine. Adv Respir Med. 2023;91(5):445–63.
Sax DR, Mark DG, Rana JS, Collins SP, Huang J, Reed ME. Risk adjusted 30-day mortality and serious adverse event rates among a large, multi-center cohort of emergency department patients with acute heart failure. J Am Coll Emerg Physicians Open. 2022;3(3):e12742.
Amatya Y, Russell FM, Rijal S, Adhikari S, Nti B, House DR. Bedside lung ultrasound for the diagnosis of pneumonia in children presenting to an emergency department in a resource-limited setting. Int J Emerg Med. 2023;16(1):2.
Baid H, Vempalli N, Kumar S, Arora P, Walia R, Chauhan U, et al. Point of care ultrasound as initial diagnostic tool in acute dyspnea patients in the emergency department of a tertiary care center: diagnostic accuracy study. Int J Emerg Med. 2022;15(1):27.
House DR, Amatya Y, Nti B, Russell FM. Impact of bedside lung ultrasound on physician clinical decision-making in an emergency department in Nepal. Int J Emerg Med. 2020;13(1):14.
Lichtenstein DA. Lung ultrasound in the critically ill. Ann Intensive Care. 2014;4(1):1.
Dong LJ, Li J, Liu W, Ankaerjiang AW, Li B, Chen YT, et al. Diagnostic efficacy of lung ultrasound in cardiogenic pulmonary edema: a systematic review and meta-analysis. Eur Rev Med Pharmacol Sci. 2023;27(15):6947–55.
Öhman J, Harjola VP, Karjalainen P, Lassus J. Rapid cardiothoracic ultrasound protocol for diagnosis of acute heart failure in the emergency department. Eur J Emerg Med. 2019;26(2):112–7.
Imanishi J, Maeda T, Ujiro S, Masuda M, Kusakabe Y, Takemoto M, et al. Association between B-lines on lung ultrasound, invasive haemodynamics, and prognosis in acute heart failure patients. Eur Heart J Acute Cardiovasc Care. 2023;12(2):115–23.
Dubón-Peralta EE, Lorenzo-Villalba N, García-Klepzig JL, Andrès E, Méndez-Bailon M. Prognostic value of B lines detected with lung ultrasound in acute heart failure. A systematic review. J Clin Ultrasound. 2022;50(2):273–83.
Apiratwarakul K, Cheung LW, Pearkao C, Gaysonsiri D, Ienghong K. Smart Emergency Call Point Enhancing Emergency Medical services on University campuses. Prehosp Disaster Med. 2024;39(1):32–6.
Apiratwarakul K, Cheung LW, Ienghong K. Impact of Smart glasses on Patient Care Time in Emergency Medical services Ambulance. Prehosp Disaster Med. 2023;38(6):735–9.
Ienghong K, Cheung LW, Wongwan P, Apiratwarakul K. Smart glasses to facilitate Ultrasound guided Peripheral Intravenous Access in the Simulation setting for Thai Emergency Medical Service providers. J Multidiscip Healthc. 2023;16:2201–6.
Gohar E, Herling A, Mazuz M, Tsaban G, Gat T, Kobal S, et al. Artificial Intelligence (AI) versus POCUS Expert: a validation study of three automatic AI-Based, Real-Time, Hemodynamic Echocardiographic Assessment Tools. J Clin Med. 2023;12(4):1352.
Gottlieb M, Patel D, Viars M, Tsintolas J, Peksa GD, Bailitz J. Comparison of artificial intelligence versus real-time physician assessment of pulmonary edema with lung ultrasound. Am J Emerg Med. 2023;70:109–12.
Pare JR, Gjesteby LA, Tonelli M, Leo MM, Muruganandan KM, Choudhary G, et al. Transfer learning-based B-Line Assessment of Lung Ultrasound for Acute Heart failure. Ultrasound Med Biol. 2024;50(6):825–32.
Mika S, Gola W, Gil-Mika M, Wilk M, Misiołek H. Overview of artificial intelligence in point-of-care ultrasound. New horizons for respiratory system diagnoses. Anaesthesiol Intensive Ther. 2024;56(1):1–8.
Tariq MA, Malik MK, Khalid Z, Asrar A. Door-to-diuretic time and short-term outcomes in Acute Heart failure. A systematic review and Meta-analysis. Crit Pathw Cardiol. 2024. https://doi.org/10.1097/HPC.0000000000000362
Sampaio Rodrigues T, Garcia Quarto LJ, Nogueira SC, Theuerle JD, Farouque O, et al. Door-to-diuretic time and mortality in patients with acute heart failure: a systematic review and meta-analysis. Am Heart J. 2024;269:205–9.
Dzikowicz DJ, Pokhrel Bhattarai S, Ng J, Zemanek AL, Carey MG. Delays in door-to-Diuretic Time and 1-Year mortality among patients with heart failure. J Cardiovasc Nurs. 2022;37(5):410–7.
Mebazaa A, Yilmaz MB, Levy P, Ponikowski P, Peacock WF, Laribi S, et al. Recommendations on pre-hospital & early hospital management of acute heart failure: a consensus paper from the Heart Failure Association of the European Society of Cardiology, the European Society of Emergency Medicine and the Society of Academic Emergency Medicine. Eur J Heart Fail. 2015;17(6):544–58.
Kane SP, Sample Size Calculator. 2024. ClinCalc: https://clincalc.com/stats/samplesize.aspx. Accessed August 22, 2024.
Choi YA, Jung JY, Park JW, Lee MS, Kim TK, Lee SGW, et al. Association between focused cardiac ultrasound and time to furosemide administration in acute heart failure. Am J Emerg Med. 2022;59:156–61.
Matsue Y, Damman K, Voors AA, Kagiyama N, Yamaguchi T, Kuroda S, et al. Time-to-Furosemide treatment and mortality in patients hospitalized with Acute Heart failure. J Am Coll Cardiol. 2017;69(25):3042–51.
Santus P, Radovanovic D, Saad M, Zilianti C, Coppola S, Chiumello DA, et al. Acute dyspnea in the emergency department: a clinical review. Intern Emerg Med. 2023;18(5):1491–507.
Gargani L. Ultrasound of the lungs: more than a room with a view. Heart Fail Clin. 2019;15(2):297–303.
Johannessen Ø, Claggett B, Lewis EF, Groarke JD, Swamy V, Lindner M, et al. A-lines and B-lines in patients with acute heart failure. Eur Heart J Acute Cardiovasc Care. 2021;10(8):909–17.
Seibel A, Zechner PM, Berghold A, Holter M, Braß P, Michels G, et al. B-Lines for the assessment of extravascular lung water: just focused or semi-quantitative? Acta Anaesthesiol Scand. 2020;64(7):953–60.
Russell FM, Ehrman RR, Barton A, Sarmiento E, Ottenhoff JE, Nti BK. B-line quantification: comparing learners novice to lung ultrasound assisted by machine artificial intelligence technology to expert review. Ultrasound J. 2021;13(1):33.
Baloescu C, Toporek G, Kim S, McNamara K, Liu R, Shaw MM, et al. Automated Lung Ultrasound B-Line Assessment using a deep learning algorithm. IEEE Trans Ultrason Ferroelectr Freq Control. 2020;67(11):2312–20.
Ienghong K, Cheung LW, Tiamkao S, Bhudhisawasdi V, Apiratwarakul K. The impact of Prehospital Point of Care Ultrasounds on Emergency patients length of stay in Thailand. J Multidiscip Healthc. 2023;16:219–26.
Acknowledgements
The authors would like to express their sincere gratitude to Josh Macknick for serving as an English consultant.
Funding
This study was funded by the Research and Graduate Studies Department, Khon Kaen University, Thailand.
Author information
Authors and Affiliations
Contributions
KI, KA, LWC, SC participated in planning the study design. KI and KA took part in the data collection. KI and KA analyzed the data and drafted the first version of the manuscript. LWC helped with statistical testing and revised the manuscript. KI, SC, and KA made revised the manuscript. KI completed the final manuscript. All authors agreed on the final version.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study and its protocol were authorized by the Khon Kaen University Ethics Committee for Human Research, in accordance with the Declaration of Helsinki and the ICH Good Clinical Practice Guidelines (HE671458). The requirement for informed consent was waived.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Ienghong, K., Cheung, L.W., Chanthawatthanarak, S. et al. Automatic B-lines: a tool for minimizing time to diuretic administration in pulmonary edema patients in the emergency department of a developing country. Int J Emerg Med 17, 183 (2024). https://doi.org/10.1186/s12245-024-00776-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12245-024-00776-1